SAT Systems was established in 1984
Sub Aquatic Technical Systems Ltd - Gas Analysis, Monitoring and Control - Data Logging and Communications offering Diving and Hyperbaric Solutions

Tuesday, March 19, 7:44 am

Series-5 *New Range of instrumentation

SAT-5
Air-5

Partial Pressure Oxygen Monitor (mA)

Description

The SAT-5 Oxygen partial pressure monitor uses an electro-chemical oxygen sensor connected to a 4-20mA transmitter with local display that is located inside the pressure compartment to be monitored. The transmitter connects to the surface display unit using a single twisted screened pair cable.

The monitor will display oxygen levels are in the range of 0 – 2.000 bar with one millibar resolution. The monitor has a history page that will display a graph of oxygen partial pressure against time with selectable time spans of between 20 minutes and 24 hours.

An internal “black box” recorder stores 12 months’ worth of data including user interactions on a micro SD card; this feature provides a useful audit trail.

Ethernet and isolated 4-20mA output connections are fitted as standard to interface the instrument with data visualisation and storage software.

The SAT-5 instruments are fully configurable from their touch screen user interface. Important settings are password protected to prevent unauthorised adjustment.

This instrument is suitable for remote monitoring of oxygen partial pressure typically in diving bells and SPHLs where the sensor is located a long distance from the primary display.

 

Specifications

Size / Weight:

Property Value
Length 210 mm
Width 106 mm
Height 128 mm
Weight 1.3 kg

Temperature Range:

Property Value
Operating 0 – 50 ºC
Storage -20 – 60 ºC

Other:

Property Value
RoHS Compliant Yes

 

Electrical Specifications

Mains Power Input:

Property Value
Voltage 100 – 240 V
Frequency 50 / 60 Hz
Power 5 W

Optional DC Power Input:

Property Value
Voltage 10 – 36 V
Power 5 W

Relays:

Property Value
Maximum voltage 250 VAC
Maximum current 6 A
Volt-free contacts Yes

4-20mA Output:

Property Value
Maximum load 600 Ω
Drive voltage 15 V
Galvanic Isolation Yes
Isolation voltage 400 Vrms

Alarms:

Property Value
Loudness 76 dB @ 1m

Sensor Specifications

Chemical oxygen sensor:

Property Value
Part number D-05
Lifetime ~5,000,000 mbar-hours, ~2.5years at 0.209 bar
Range 0 – 2 bar
Accuracy ± 5 mbar
Response Time T90 <14s

 

4-20mA Oxygen transmitter:

Property Value
Part number B03-35-01
Lifetime ~5,000,000 mbar-hours, ~2.5years at 0.209 bar
Range 0 – 2 bar
Accuracy ± 5 mbar
Response Time T90 <14s

 

Request Quotation
Request a Quotation

Thank you for your interest in receiving a Quotation for our instrumentation.


Below you will find three short steps to follow in order for us to produce a quotation. To enable us to deal with your request efficiently, please ensure that you supply your full Company details together with your exact requirements.

Please note: all quotations are manually generated based on the information you will supply. Details of your intended supply voltage (if applicable) will ensure we supply the correct product code and lead time.
step 1 step 2 step 3
Please register on our Helpdesk, supplying your full company details Please check the email account that you used when registering, we have sent you an email with a link to activate your account Once your account is activated - Login to our helpdesk and select Request and Quotation
Please Click Here To Register Now! If you do not receive our activation email or experience any problems registering, please Contact Us and we will manually create your account. Login To Our Helpdesk Now!

Data Sheets

Please use the form below to request data sheets for our New Series-5 Range of instrumentation.

Please Note: All data sheets are sent manually to the email address you supply - requests can take up to 48 hours to complete and are handled during our normal office hours.

Thank You


Data Sheet Request Form

Data Sheets




























S.A.T.Systems Ltd   1 Kelly Industrial Estate, Kirk Michael, Isle of Man,  IM6 1SS