The AIR-5 hyperbaric oxygen and carbon dioxide monitor is designed for monitoring the environment inside a pressurised compartment. The monitor is supplied as two components: a touch screen display module and a sensor pack with local display.
The display module has a colour touch screen that is the operator interface. All adjustments, configuration and user interaction is carried out through this graphical interface. High and low alarms can be set for oxygen, high and higher alarms can be set for carbon dioxide. All settings are password protected and changes are logged. The display module connects to the sensor pack via a pressure proof penetrator using a two twisted pair overall screened cable. This cable is terminated in an IP67 connector at the display module and through a watertight gland into the sensor pack.
The sensor pack is fitted with sensors for oxygen, carbon dioxide and ambient pressure; the sensors measure the ambient gas concentrations by diffusion through hydrophobic membranes. The sensor pack includes a local LED display and its enclosure is designed for use in a hostile environment.
The instrument records the readings and alarm status every second in non-volatile memory. The built in storage is capable of recording a year’s continuous data, and if required, a larger data storage capacity can easily be installed. The data is also available for remote recording in real time through the RS485 port.
The unit is powered from a 9-36 VDC power source. All the components are designed for use in a marine environment and the electronics meet the current European standards for electrical safety, electromagnetic emissions and immunity.
Order Code: | C08-001-009-01-*-1 |
Description: | AIR-5 Hyperbaric Oxygen and Carbon Dioxide Monitor |
Display & Control |
|
Display: | 3.5" Colour TFT with LED backlight |
Touch Screen: | Resistive |
Internal Data Storage: | 1 year continuous recording logged every second (upgradable) |
External Data Connection: | Isolated RS485 |
Power Requirements: | 9 to 36 VDC (5W) |
Operating Temperature: | 0 to +50 ºC |
Storage Temperature: | -20 to +60 ºC |
Oxygen Sensor |
|
Range: | 0 to 2 bar oxygen |
Accuracy: | ±0.5% of Full scale |
Response Time: | T90 < 14 Seconds |
Resolution: | 1 mbar |
Expected Life: | 5,000 bar 02 hours |
Carbon Dioxide Sensor |
|
Range: | 0 to 50 mbar carbon dioxide |
Accuracy: | +/- 5% of reading |
Response Time: | T90 < 30 Seconds |
Resolution: | 0.1 mbar |
Expected Life: | > 5 Years |
Pressure Sensor |
|
Range: | 0 to 10 bar(a) |
Accuracy: | +/- 0.15% of full scale |
Response Time: | 0.1 Second |
Resolution: | 1 mbar |
Environmental Specifications |
|
Pressure Compensation: | 0.7 to 6 bar(g) |
Operating Temperature: | 0 to +50 ºC |
Storage Temperature: | -20 to +60 ºC |
Built in Display |
|
Type: | White on Blue OLED |
Size: | 128 x 64 pixel |
45 x 29 mm |
Thank you for your interest in receiving a Quotation for our instrumentation. |
||
Please register on our Helpdesk, supplying your full company details | Please check the email account that you used when registering, we have sent you an email with a link to activate your account | Once your account is activated - Login to our helpdesk and select Request and Quotation |
Please Click Here To Register Now! | If you do not receive our activation email or experience any problems registering, please Contact Us and we will manually create your account. | Login To Our Helpdesk Now! |
Please use the form below to request data sheets for our New Series-5 Range of instrumentation.
Please Note: All data sheets are sent manually to the email address you supply - requests can take up to 48 hours to complete and are handled during our normal office hours.
Thank You