The Air-5 oxygen analyser & hot water temperature monitor is designed to monitor the gas supply and hot water feed to a diver in the water. The analyser is supplied as two components: a touch screen display module, either panel mount or surface mount, and a sensor pack.
The display module has a colour touch screen which is the operator interface. All adjustments, configuration and user interaction is carried out through this graphical interface. High and low alarms can be set for both the oxygen and hot water supply.
The sensor pack includes a sample flow controller and an electronic flow meter with flow failure alarm, which can be located in a convenient place close to the diver’s breathing gas supply pipework where the sample is teed off and connected directly to the inlet of the sensor pack. The temperature sensor should be installed in a suitable fitting that is part of the diver’s hot water supply manifold.
The display module connects to the sensor pack using a 2 twisted pair overall screened cable. This cable is terminated in either a D-9 connector at the panel mount display module, or an IP67 connector at the surface mount display and through a watertight gland into the sensor pack.
The instrument records the readings and alarm status every second in non-volatile memory. The built in storage is capable of recording a year’s continuous data, and if required, a larger data storage capacity can easily be installed.
The unit is powered from a 9-36 VDC power source.
All the components are designed for use in a marine environment and the electronics meet the current European standards for electrical safety, electromagnetic emissions and immunity.
Model : AIR-5 Oxygen and Temperature
Part Number: C08-002-002-04-*-*
Range—Oxygen : 0 to 100% Oxygen
Range—Temperature : Nominal 0-100°C (others available)
Accuracy—Oxygen : 1% of Full Scale
Accuracy—Temperature : 1% of Full Scale
Resolution : 0.1°C
Response Time—Oxygen : T90 < 14 Seconds
Response Time—Temperature : Depends on sensor selection.
Housing : 110*140*60 mm (Width, Length, Depth)
Expected Oxygen Sensor Life : 500,000 Vol.% hours
Power Requirements : 10 to 36VDC(5W)
Operating Temperature : 0 to +50 °C
Storage Temperature : -20 to +60 ºC
Weight : 0.9Kg
RoHS Compliant : Yes
Thank you for your interest in receiving a Quotation for our instrumentation. |
||
Please register on our Helpdesk, supplying your full company details | Please check the email account that you used when registering, we have sent you an email with a link to activate your account | Once your account is activated - Login to our helpdesk and select Request and Quotation |
Please Click Here To Register Now! | If you do not receive our activation email or experience any problems registering, please Contact Us and we will manually create your account. | Login To Our Helpdesk Now! |
Please use the form below to request data sheets for our New Series-5 Range of instrumentation.
Please Note: All data sheets are sent manually to the email address you supply - requests can take up to 48 hours to complete and are handled during our normal office hours.
Thank You